Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15262
Brand Name:Stryker OtisMed Shapematch Cutting Guides
Model:Catalogue Numbers: TR3100-L, TR3100-R
Affected:All lot numbers
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Howmedica Osteonics
Issue:The reason for this Hazard Alert is multifactorial:

1/ The surgical protocol for the ShapeMatch Cutting Guides did not provide sufficient information on:
The position of the cutting guides,
The need for osteophyte/soft tissue removal,
The requirement for proper axial drilling and
The need to avoid angulation and mal-alignment.

2/ The approved pre-operative plan did not provide sufficient information regarding the need for osteophyte removal.

3/ The values displayed via the OtisMed.net web portal were inconsistent with the pre-operative planning values used to manufacture some of the ShapeMatch Cutting Guides.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:2/9/2013