Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15242 |
| Brand Name: | Siemens EasyLink Data Management System |
| Model: | Catalogue: 10444799, 10444800, 10487219 |
| Affected: | Version V4.0 and V5.0 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Under certain conditions the system may not perform as intended, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay / omission of result transmission to the LIS |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/8/2013 |