Detail of Recall

Type of Product:Device
Medsafe Reference:15226
Brand Name:GE Discovery & Optima Magnetic Resonance Imaging Systems
Model:All Discovery MR450, Discovery MR750, and Optima MR450w products
Affected:All serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Medical Systems LLC
Issue:The yellow foam strip used between the end bell covers and patient bore of certain GE MR Products may extend past the surface of the covers. The foam itself is not hazardous. However, any fluid or body substance absorbed by the extended foam may come into contact with patients and act as a fomite potentially transferring infectious agents from one patient to another.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/8/2013