Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15219
Brand Name:Ascension PyroTitan Humeral Resurfacing Arthroplasty components
Model:12 models
Affected:All lots
Software version:
Recalling Organisation:LMT Surgical
9 Business Connection
2 Kalmia Street
Ellerslie Greenlane
Auckland 1544
Contact Information:
Manufacturer:Ascension Orthopedics Inc
Issue:The manufacturer, Ascension Orthopaedics, has received complaints of implant fractures that were observed post-operatively and required revision surgeries. Most confirmed cases of implant breakage have occurred as a result of excessive loading and within a year of being implanted. Excessive loads placed on the implant through high impact activities or sudden trauma can damage an artificial joint, particularly in the presence of poor bone stock. High impact activity may cause loosening or fracture of the implant. The breakage can result in glenohumeral joint pain and possible damage to the surrounding tissues.
Recall action type:Alert
Recall action:Manufacturer to issue advice regarding use
Level of recall:Healthcare Professional
Date Commenced:30/8/2013
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /