Detail of Recall

Type of Product:Device
Medsafe Reference:15191
Brand Name:GE Tec 6 & Tec 6 Plus Desflurane Vaporizers
Model:Models: Tec 6 & Tec 6 Plus
Affected:All serial numbers
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Healthcare Finland Oy
Issue:The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to the atmosphere and a reduction in fresh gas delivered to the breathing system.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/8/2013