Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15191 |
| Brand Name: | GE Tec 6 & Tec 6 Plus Desflurane Vaporizers |
| Model: | Models: Tec 6 & Tec 6 Plus |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Healthcare Finland Oy |
| Issue: | The user manual pre-operative check, low pressure leak test may not detect the full range of leaks from seal wear degradation in the vaporizers. If the leak is not detected, it will result in fresh gas leaking to the atmosphere and a reduction in fresh gas delivered to the breathing system. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/8/2013 |