Detail of Recall

Type of Product:Device
Medsafe Reference:15144
Brand Name:Ortho-Clinical Diagnostics VITROS Chemistry Products MicroSlides
Model:Multiple catalogue numbers
Affected:Expiry dates to 22 March 2015
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics c/- Johnson & Johnson
13A Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 800 705
Manufacturer:Ortho-Clinical Diagnostics Inc
Issue:The VITROS MicroSlide Cartridge is made up of the cartridge body and an Anti-Backup Platen(ABP). The purpose of the ABP is to keep the stacks of slides in the correct position for dispensing. Ortho Clinical Diagnostics have received reports associated with ABP failing to advance, jamming or failing to stay locked in place within the cartridge. When this occurs, the slide is not properly dispensed from the cartridge since the stack of slides is not positioned correctly.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/8/2013