Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15098 |
| Brand Name: | ASP Sterrad 100NX Cassette |
| Model: | Part Number: 10144 |
| Affected: | Expiry date prior to 1-Nov-2014 |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Advanced Sterilization Products Inc |
| Issue: | The manufacturer, ASP has determined it does not have adequate data to support the effectiveness of the packaging's Leak Indicator for the entire duration of the STERRAD® 100NX® System cassette's 15-month shelf life. The likelihood of a leak is extremely low, and there have been no reported cases of a leak where the indicator was not effective in identifying the leak. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 2/8/2013 |