Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15095 |
| Brand Name: | ArjoHuntleigh Contoura 380 & 480 Beds |
| Model: | Models: 380 & 480 |
| Affected: | Manufactured between 12 January 2011 to 31 December 2011 |
| Software version: | |
| Recalling Organisation: | Arjo New Zealand Limited 34 Vestey Drive Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 000 151 |
| Manufacturer: | ArjoHuntleigh AB |
| Issue: | During investigations by ArjoHuntleigh it has been determined that the DIN connector used on these beds has been compromised during the cable manufacturing process. This could result in the beds performing uncommanded movements. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/8/2013 |