| Type of Product: | Device |
| Medsafe Reference: | 15073 |
| Brand Name: | Bio-Rad Seraclone Anti-S |
| Model: | Catalogue: 808052 |
| Affected: | Lot Numbers: 2223041, 2244061 |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Bio-Rad Laboratories Inc |
| Issue: | Stability data for Seraclone Anti-S showed a reactivity decrease after approximately 12 months shelf life for the lot numbers listed above. Weak or false negative results might be possible in testing. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/7/2013 |