Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:15061
Brand Name:Siemens Advia Centaur Folate Assay
Model:
Affected:22288218, 22296218, 22907218, 22289218, 22297218, 22535218, 22298218 and 22299218
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed that a portion of Folate readyPacks in kit lots are exhibiting calibration failures and/ or significant negative shifts in quality control (QC) and patient results. The negative bias could potentially lead to additiional vitamin supplementation or dietary modification.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/7/2013