Detail of Recall

Type of Product:Device
Medsafe Reference:15059
Brand Name:Ortho Clinical Diagnostics BioVue System Cassettes
Model:BioVue ABD confirmation Cassette, VioVue Rh/K Cassette, BioVue ABO-DD Grouping Cassette, BioVue ABO-Rh Grouping Cassette, BioVue Poly/Neutral Cassette
Affected:Lots: 6904485 and 707135
Software version:
Recalling Organisation:Ortho-Clinical Diagnostics c/- Johnson & Johnson
13A Gabador Place
Mt Wellington
AUCKLAND 1060
Contact Information:0800 800 705
Manufacturer:Ortho-Clinical Diagnostics European Support Centre
Issue:Ortho Clinical Diagnostics (OCD) has identified isolated occurrences of improperly positioned cassette labels for the products specified. The use of an affected multi-reagent cassette may lead to false negative or false positive results causing a potential misclassification of the patient or donor blood groups or incorrect antibody detection results. There is no risk associated with the use of an affected single-reagent cassette since all wells contain the same reagent.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/8/2013