Detail of Recall

Type of Product:Device
Medsafe Reference:15039
Brand Name:Philips Brilliance iCT and iCT SP
Model:
Affected:
Software version:up to an inc,uding version 3.2.5
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:When using 80 kVp with intravenous contrast in head and neck studies with U-filters (UA, UB, UC) a vascular artifact that resembles thrombus may appear on the image. There is a risk of misdiagnosis which can lead to unnecessary treatment of a patient.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:25/7/2013