Detail of Recall

Type of Product:Device
Medsafe Reference:15027
Brand Name:Philips Briiliance (BR) CT 16-Slice, BR CT 10-Slice Water, BR CT 6-Slice Water, MX8000 Quad, MX8000 Dual, MX8000 Dual EXP, MX8000 IDT and MX8000 IDT 10
Model:All models
Affected:All serial numbers
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The CT system phantom supplied with the above-listed scanners has a special pin by which the customer can check the Low Contract (LC) detectability for quality assurance. This pinconsists of a Nylon (Aculon) body with six smaller Lexan pins of 3mm, 4mm, 5mm, 6mm, 7mm and 8mm diameters, which have contrast difference of approximately 1% from Aculon.
The LC test is performed by visually inspecting the sectionand selecting the smallest detectable Lexan pin for a predefined scan protocol as defined in the instructions for use. The manufacturer-recommended procedure is subjective in nature and needs to be performed on several scans and by different people to be reliable.
The pin visibility deteriorates over time because of slow water absorption. As the pin visibility deteriorates the test becomes unreliable and inaccurate.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/7/2013