Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 15021 |
| Brand Name: | Baxter Clearlink Solution Sets |
| Model: | FNC3110 |
| Affected: | SR13A04021, SR13A02041, SR13A07032, SR13B14077 |
| Software version: | |
| Recalling Organisation: | Baxter Healthcare Ltd 33 Vestey Drive Mt Wellington AUCKLAND |
| Contact Information: | 0800 229 837 |
| Manufacturer: | Baxter Healthcare Corporation |
| Issue: | Manufacturer received feedback from a customer that an IV tube could not be loaded into a infusion pump because the side clamp was not correctly assembled. Manufacturer confirmed that the complaint sample and confirmed side clamp assembled in the reverse direction. The incorrect assembly would lead to incorrect direction of infusion such that blood may withdrawn froma patient and result in acute blood loss and/ or delay in treatment of life sustaining medication as worst case scenario. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Wholesale |
| Date Commenced: | 29/7/2013 |