Detail of Recall

Type of Product:Device
Medsafe Reference:15014
Brand Name:Radiometer pCO2 membrane (D788) for ABL7/8xx analyzers
Model:Catalogue: 942-063
Affected:all lots
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:Possible negative bias on pCO2 result on both blood and QC after membrane replacement. This bias can be seen after the replacement of the membrane.
Update: The temporary countermeasures, regarding the additional requirement for manually narrowing the QC ranges for pCO2 and also performing extra QC measurements as communicated in July 2013 are no longer required and may be revoked upon installation of the improved sealing ring on the pH electrodes in ABL700 and ABL800 Series Analyzers.
01-02-2017: This notification is to inform you that your Radiometer representative has now installed the improved sealing ring for the pH electrode in your ABL700 and/or ABL800 series analysers.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/11/2016