Detail of Recall

Type of Product:Device
Medsafe Reference:15011
Brand Name:Medtronic Navigation Framelink Software
Model:9734047
Affected:all lots of version 5.4
Software version:5.4
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:Manufacturer has identified that the use of high resolution exams (greater than 256 x 256) under certain use conditions may cause the Look-Ahead view labels to show user selected depths (displayed) , but the corresponding images are shown at a depth twice the selected distance. Under these conditions visulaization of the planned trajectory could show the image at the incorrect depth. Further internal investigation discovered that the use of high resolution exams (256 x 256) under certain use conditions may cause a similar discrepancy between the exam image and the displayed measurement or overlay.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:17/7/2013