Detail of Recall

Type of Product:Device
Medsafe Reference:14991
Brand Name:Elekta Monaco Radiation Treatment Planning System
Model:
Affected:
Software version:3.100 and later
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:Sample couch parameter files are supplied with Monaco release 3.100 and later. The user can then access these files from the Treatment Couch Library.By clicking on a particular couch file, the individual components of the couch can be selected and their densities edited. The user can then use the "Save as Treatment Couch" to enter a unique name for the modified couch parameter file.
When sample couch parameter files are edited, the densities altered, and "Save" is selected, the sample files will be updated with these new, user defined densities. Subsequently, when a software upgrade is loaded, the modified couch densities in these samples will be overwrittem with the default value of 1.000.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/7/2013