Database updated: 5 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14957
Brand Name:Medtronic MiniMed Insulin Infusion Pump Reservoirs
Model:Models: MMT326A & MMT-332A
Affected:Lots: H8442973, H8469703, H8496561, H8517079, H8463297, H8492449, H8510440, H8539013, H8467888, H8494645, H8512826
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic MiniMed
Issue:Affected lots of insulin reservoirs have an increased risk of leaking. A leak in the reservoir may result in delivery of less insulin than intended and, if there is an occlusion in the infusion set, the pump may not alarm.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Public Recall
Date Commenced:5/7/2013
Additional Information:www.medsafe.govt.nz/hot/RecallActionNoticesNew/14957.asp
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /