Detail of Recall

Type of Product:Device
Medsafe Reference:14944
Brand Name:Elekta Digital Accelerators
Model:All Digital Accelerators with Electrons and Beam Modulator, MLCi/MLCi2, Agility or Asymmetric Heads
Affected:All serial numbers
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Issue:Elekta is aware that some Digital Accelerators have increased electron applicator auto-tracking settings in the field, to values greater than the factory-set defaults.
Increasing auto-tracking values from factory-set defaults goes against Elekta recommendations and can result in the system becoming non-compliant with IEC standards, in particular with the safety standard IEC 60601-2-1.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/7/2013