Detail of Recall

Type of Product:Device
Medsafe Reference:14926
Brand Name:Siemens Dimension Vista 500 & 1500 analysers
Model:10488224 & 1044802
Affected:
Software version:3.5.1 or 3.6
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:Manufacturer has confirmed that, under specific rare conditions which require a system reset, there is a potential for a sample to be dispensed into an aliquot well that has previously been used for a QC or calibrator from a Vista vial. If two samples are dispensed into the same aliquot well, the first sample will always be QC or calibrator from a Vista vial. Patient sample will never be dispensed into an aliquot well that contains another patient sample.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/6/2013