Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14926 |
| Brand Name: | Siemens Dimension Vista 500 & 1500 analysers |
| Model: | 10488224 & 1044802 |
| Affected: | |
| Software version: | 3.5.1 or 3.6 |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | Manufacturer has confirmed that, under specific rare conditions which require a system reset, there is a potential for a sample to be dispensed into an aliquot well that has previously been used for a QC or calibrator from a Vista vial. If two samples are dispensed into the same aliquot well, the first sample will always be QC or calibrator from a Vista vial. Patient sample will never be dispensed into an aliquot well that contains another patient sample. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/6/2013 |