Detail of Recall

Type of Product:Device
Medsafe Reference:14917
Brand Name:BioCSL Phenocell C Reagent, 0.8%
Model:Catalogue: 96530201
Affected:Lots: 9653052, 9653053, 9563054
Software version:
Recalling Organisation:Seqirus Pty Ltd
63 Poplar Road
Parkville
VIC 3052
Australia
Contact Information:
Manufacturer:Seqirus Pty Ltd
Brand Name:BioCSL Phenocell C Reagent, 3%
Model:Catalogue Number: 86640201
Affected:Lots: 8664052, 8664053, 8664054
Software version:
Recalling Organisation:Seqirus Pty Ltd
63 Poplar Road
Parkville
VIC 3052
Australia
Contact Information:
Manufacturer:Seqirus Pty Ltd
Issue:The Antigen Composition Sheet supplied with the product listed Cell 5 as Lua negative. However, a customer complaint was received from the IMVS Adelaide Red Cell Reference Laboratory that they typed Cell 5 as Lua positive. Confirmatory testing performed by bioCSL in response to the customer complaint has confirmed that Cell 5 did in fact type as Lua positive.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/7/2013