Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14891
Brand Name:Synthes Transpedicular Schanz Screw
Model:Part Numbers: 296.690, 496.711, 496.712, 496.721
Affected:Various lots
Software version:
Recalling Organisation:Synthes New Zealand Ltd
507 Mt Wellington Highway
Mt Wellington
Auckland 1060
Contact Information:0800 657 734
Manufacturer:Synthes GmbH
Issue:Synthes received several complaints that the clamps could not be connected with the Transpedicular Schanz Screw, because in some cases, the shaft of the screw is too thick.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/7/2013