Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14888
Brand Name:Respironics V60 Ventilators
Model:Model: V60
Affected:Manufactured between Mar-2009 and Apr-2013
Software version:
Recalling Organisation:Care Medical Ltd
Contact Information:0800 333 444
Manufacturer:Respironics California LLC
Issue: If a component fails on the Power Management Board Assembly, it may cause ventilator support to be lost with potentially no audible alarm from the ventilator.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/7/2013
Hide menus
Show menus