Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14888 |
| Brand Name: | Respironics V60 Ventilators |
| Model: | Model: V60 |
| Affected: | Manufactured between Mar-2009 and Apr-2013 |
| Software version: | |
| Recalling Organisation: | Care Medical Ltd |
| Contact Information: | 0800 333 444 |
| Manufacturer: | Respironics California LLC |
| Issue: | If a component fails on the Power Management Board Assembly, it may cause ventilator support to be lost with potentially no audible alarm from the ventilator. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/7/2013 |