Detail of Recall

Type of Product:Device
Medsafe Reference:14847
Brand Name:Philips Heartstart MRx Monitor/Defibrillator
Model:Models: M3535A, M3536A, M3536J, M3536M, M3536MC, M3536M2, M3536M4, M3536M5, M3536M6
Affected:Serial numbers within the range US00100100 to US00567299
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The Philips HeartStart MRx Monitor/Defibrillator can deliver a non- synchronised shock in the synchronised cardioversion mode. This can occur if the user rotates the Therapy Knob while simultaneously pressing the Sync button, then charges the MRx and presses the shock button.
The MRx will still display R wave markers as if normally operating in synchronised cardioversion mode. The "Sync" Label will still be seen on the display. When the MRx is charged and the shock button pressed, the shock will be immediately delivered and may not be synchronised to the R wave.
In addition, the label affixed to the top of the MRx device is not consistent with the MRx Instructions for Use on synchronised cardioversion.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2013