Detail of Recall

Type of Product:Device
Medsafe Reference:14837
Brand Name:Mesa EZTest Biological Indicators
Model:Models: EZS/6 & EZS5/25
Affected:Lots: S-445 & S-447
Software version:
Recalling Organisation:W M Bamford & Co Ltd
12 Victoria Street
LOWER HUTT
Contact Information:04 576 2100
Manufacturer:Altrax Group Ltd
Issue:The manufacturer, Mesa, recently determined that the read out time is not in specification with the label claims on these lots. There is potential for delayed or incomplete colour change (false fail) at the indicated incubation time of 10 hours.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/6/2013