Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14828 |
| Brand Name: | Beckman Coulter Anti-Mullerian Hormone (AMH) Gen II ELISA |
| Model: | Part Number: A79765 |
| Affected: | All lots before and including 326119 |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | Undiluted patient samples masured with the AMH Gen II ELISA kit may generate results that are lower than expected, due to interference from complement. This may lead to test results for neat AMH patients samples being lower than expected. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/6/2013 |