Type of Product:Device
Medsafe Reference:14800
Brand Name:Bio-Rad- IH-1000 Analyser
Model:Model: 001000
Affected:All software versions
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:Under specific conditions inconsistent grading between the results in the well/card and the results returned by the IH-1000 may be observed.
this issue might lead to a wrong result released to the host if all the 3
following conditions are met:
1. A result is returned as negative while the reaction is positive, DP, or requires a human interpretation and,
2. This negative result is consistent with the global interpretation of the test and,
3. The system is configured without the second reading function activated for all tests results.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:13/6/2013