Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14791 |
| Brand Name: | CareFusion Alaris GP, GP Guardrails, GP Plus & GP Plus Guardrails Infusion Pumps |
| Model: | Models: 80263UN01, 80263UN01-G, 9002MED01, 9002MED01-G |
| Affected: | Pumps manufactured between Dec 2009 and Jan 2012 |
| Software version: | |
| Recalling Organisation: | CareFusion New Zealand 313 Limited Unit 4 / 110 Mays Road Onehunga AUCKLAND 1061 |
| Contact Information: | 0508 422 734 |
| Manufacturer: | CareFusion Switzerland 317 Sarl |
| Issue: | The manufacturer has identified an increased occurrence of stepper motor stalls with selected models of Alaris GP infusion pump. It has been determined that these stalls are due to the stepper motor's front and rear bearings which may not perform as designed after a period of use causing the pump to stop infusing. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 11/6/2013 |