Detail of Recall

Type of Product:Device
Medsafe Reference:14762
Brand Name:Elekta Monaco Radiation Treatment Planning System
Model:
Affected:All versions
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:Patients can be scanned either Head First or Feet First and these CTs imported into Monaco for radiation treatment planning. The Monaco CT Sim workflow gives an option to prepare a Feet First plan, but Monaco IMRT workflow does not. Although a patient can be scanned Feet First and these Feet First CTs imported into Monaco, the user could incorrectly assume that the intensity-modulated radiation therapy (IMRT) plan is Feet First as well. When the patient is treated, they could be incorrectly positioned in the Feet First position which will not match the plan. If the patient is positioned for treatment in the Feet First position, the IMRT treatment plan will not match what is delivered, potentially resulting in a significant beam geometry error.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/6/2013