Detail of Recall

Type of Product:Device
Medsafe Reference:14760
Brand Name:Stryker Sonopet Console
Model:Catalogue: 5450852000
Affected:Serial Numbers: 1134000114, 1134000124, 1134000134, 1134000144, 1134000184, 1134000204, 1209400174
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Medtech K.K.
Issue:Stryker has received several reports that the Irrigation Pump was not functioning. In the case where the irrigation pump fails, saline would not be provided to the tip. As a result the cooling function would be lost and could potentially cause the tip to heat up potentially resulting in tissue burns to the patient
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/6/2013