Type of Product:Device
Medsafe Reference:14741
Brand Name:Oxoid Antimicrobial Susceptibility Testing Disc, Cefovecin
Model:Model: CT1929B
Affected:Lot: 1257665
Software version:
Recalling Organisation:Oxoid NZ
3 Atlas Place
Mairangi Bay
AUCKLAND 1333
Contact Information:09 980 6700
Manufacturer:Thermo Fisher Scientific Microbiology Basingstoke
Issue:A technical investigation has indicated that cartridges of this lot of Oxoid Antimicrobial Susceptibility Testing Disc CVN30 (Cefovecin), may contain discs with insufficient amount of antibiotic. Continued use of this lot could result in false indication of resistance to Cefovecin.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/5/2013