Detail of Recall

Type of Product:Device
Medsafe Reference:14740
Brand Name:Philips Precedence SPECT/CT System utilising Brilliance CT
Model:Product Code(s): 882350, 882351
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Radiation Oncology Systems
Brand Name:Philips Brilliance CT Systems
Model:Product Code(s): 728256, 728251, 728246, 728240, 728235, 728231, 728243, 728244, 728306, 728311
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Brand Name:Philips Gemini PET/CT Combination utilising Brilliance CT
Model:Product Code(s): 882160, 882300, 882390, 882400, 882410, 882412, 882470, 882473, 882471, 882476
Affected:
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems
Issue:The stated tolerance for CTDIvol in the Instructions for Use Manual did not accurately represent all systems. On some systems the measured CTDIvol value may fall outside of these stated tolerances
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/5/2013