Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14688 |
| Brand Name: | Zimmer Air Dermatome |
| Model: | Part number: 00-8801-001-00 |
| Affected: | Serial Numbers: 109445, 111058, 111059, 111067, 111101, 111102, 111122 |
| Software version: | |
| Recalling Organisation: | Zimmer Biomet New Zealand Company 210 Khyber Pass Road Grafton Auckland |
| Contact Information: | 09 925 5200 |
| Manufacturer: | Zimmer Surgical, Inc. |
| Issue: | Broken planetary gear teeth could cause the Zimmer Air Dermatome to either not operate or operate intermittently. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 20/5/2013 |