Detail of Recall

Type of Product:Device
Medsafe Reference:14664
Brand Name:Bio-Rad IH-1000 Analyser
Model:Product Ref: 001000
Affected:All software versions
Software version:
Recalling Organisation:Bio-Rad New Zealand
Level 30
48 Shortland Street
AUCKLAND 1010
Contact Information:0508 805 500
Manufacturer:Bio-Rad Laboratories Inc
Issue:An error relative to the liquid reagents management (test cell or anti-serum vials) might occur if the following conditions are present:
1. IH-1000 is in in status :ready to start".
2. The user requests the unloading of a reagent rack while the system is creating the internal batch.
3. Another is loaded immediateky afterwards in the same position.
If all three conditions are present the IH-1000 may not use the full set of iquid reagents required to perform the tests of the ongoing batch.
Due to a software failure, the system mau only use the vial palced in the first position of the re-loaded rack for all the tests of the ongoing batch.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/6/2013