Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14656 |
| Brand Name: | GE Signa Magnetic Resonance Imaging System |
| Model: | Models: 1.5T Signa HDxt, 3.0T Signa HDxt, Signa Vibrant |
| Affected: | PACS systems manufactured by Intelerad (IntelePACS 4-3-1-P350 and InteleBrowser -4-3-1-P350) |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | GE Medical Systems LLC |
| Issue: | For ClariView and Screen Save from GVTK (Gems Visualization ToolKit) Viewer, the MR Product can generate image headers with non-unique series identifiers, which when exported to some PACS systems, results in the series being associated with the incorrect study on the PACS. The intended study will not have the correct series listed and an unintended study will have it listed instead. In all cases, patient identifying annotation remains correct. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/5/2013 |