Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14638
Brand Name:Applied Medical Epix and Direct Drive Laparoscopic Graspers
Model:Models: APP-C4130, APP-C4130-S
Affected:Lot Numbers: 1145945, 1147900, 1148677, 1154637, 1159491, 1162504, 1168638, 1171631
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Applied Medical Resources Corporation
Issue:If a great enough force is applied when the ratchet trigger is activated, there is a potential for the trigger to fracture and the jaws to remain in the closed position.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:15/5/2013