Detail of Recall

Type of Product:Device
Medsafe Reference:14609
Brand Name:Medtronic NIM Flex EMG Endotracheal Tubes
Model:8229960, 8229965, 8229970, 8229975, 8229980 and 8229985
Affected:78 lot numbers
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Xomed
Issue:Tubes used to manufacture the NIM Flex Endotracheal Tube may have the potential to delaminate. Investigation has shown that such delamination can result in herniation of the inner lumen when the cuffs are over-inflated. Such herniation can contrict the inner lumen of the endotracheal tube resulting in a reduction of airflow to the patient.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/5/2013