Detail of Recall

Type of Product:Device
Medsafe Reference:14588
Brand Name:Hologic Mini C-arm
Model:InSight and InSight 2
Affected:
Software version:
Recalling Organisation:Pharmaco (NZ) Ltd
Level 1
4 Fisher Crescent
Mt Wellington
AUCKLAND 1060
Contact Information:0800 804 079
Manufacturer:Hologic Inc
Issue:Manufacturer has identified the following compliance issues:
1.The audible alarm to alert the operator when cumulative irradiation time reaches five minutes is not accurate.
2. The total cumulative irradiation time display is not accuarate.
Both issues are caused by computational error which truncates the fractions of seconds after each image and therefore the time to alarm and the time displayed are shorter by the sum of these fractions.
This non-compliance does not pose a health hazard, as the radiation exposure to a patient depends on power settings during the procedure. The system provides the user with accurate dose to the patient information by displaying Air Kerma and Cumulative Air Kerma.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/6/2013