Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14584
Brand Name:FlexStent Femoropopliteal Self-Exanding Stent System
Model:
Affected:all lots
Software version:
Recalling Organisation:Endoventure Ltd
2/1 Esmonde Road
Takapuna
Auckland
Contact Information:
Manufacturer:Flexible Stenting Solutions Inc
Issue:Manufacturer has determined that during internal package integrity testing, following transportation simulation challenge, the sterile barrrier of the pouch (inner pouch) of some units of the FlexiStent could be compromised.and to a lesser degree, the outer pouch.
To date no complaints associated with this potential mode of failure have been reported.
In the event a hole or a tear in the clear film side of the pouch of the affected product is present and both pouches are affected, loss of the sterile barrier could occur, leading to use of a non-sterile device in the procedure.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:2/5/2013