Type of Product:Device
Medsafe Reference:14572
Brand Name:GE LOGIQ S8 and LOQIQ S7 Ultrasound
Model:
Affected:with software revisions R.1.1.1 , R1.5.1 and R1.0.1, R1.0.2 and R1.0.3
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:GE Ultrasound Korea Ltd
Issue:Manufacturer has recently become aware of a potential safety issue due to the S4-10-D Ultrasound Probe used in conjunction with the LOGIQ S8 and LOGIQ S7 Ultrasound systems. It was also discovered that this issue also leads to a specification issue of the acoustic output of this probe
There is a potential for a skin burn when the flow model in the cardiac application is activated with the S4-10-D Ultrasound Probe.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/4/2013