Detail of Recall

Type of Product:Device
Medsafe Reference:14570
Brand Name:Varian RPM Respiratory Gating System
Model:H50
Affected:1.7.5 with 3D option SN's: H501056, H501080, H501944, H502289, H502300, H502301, H503567, H503596, H503935, H504201 and H504724.
Software version:
Recalling Organisation:Varian Medical Systems Australasia Pty Ltd (NZ)
Millenium Centre, Level 3, Building C, 600 Great
South Road Ellerslie
Auckland 1051
Contact Information:0800 486 732
Manufacturer:Varian Medical Systems Inc
Issue:Manufacturer has identified an anomaly with the RPM Respiratory gating System. While operating with the 6-dot marker block, this version of the RPM Gating may produce an inverted respiratory waveform in amplitude and phase-gating sessions.
The error may be reproduced if the RPM System is induced to frequently lose and re-acquire tracking of the 6-door marker block during a calibration sequence, a reference session or a single session after tracking has begun.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/4/2013