Detail of Recall

Type of Product:Device
Medsafe Reference:14569
Brand Name:Elekta Integrity 3.0
Model:
Affected:
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Limited (England)
Issue:During beam delivery, the tolerance values are used to do a check for unwanted movements. These values are known as "static tolerances" which are related to all the digital accelerator geometric movement axes except Table Column. The problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values. A fault occurs if the loaded database is incorrect
Recall action type:Recall
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/4/2013