Detail of Recall

Type of Product:Device
Medsafe Reference:14558
Brand Name:Panbio Dengue IgM Capture ELISA
Model:E-DEN01M/ E-DEN01M05
Affected:12306, 12358 and 13007
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd - Rapid Diagnostics Divison
Ground Floor, Bldg. D
4 Pacific Rise
Mount Wellington
AUCKLAND 1060
Contact Information:0800 466 951
Manufacturer:Inverness Medical Innovations Australia Pty Ltd t/a Alere
Issue:Manufacturer's QA program has determined that a reassignment of a customer aceptance range is required for three lots of product. Manufacturer received 5 complaints of invalid test runs due to results falling outside the cut-off value customer accetpance range leading to an invalid test. Analysis has indicated that the invalid test runs are due to the customer acceptance range being inappropriately assigned for these batches.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/4/2013