Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14534
Brand Name:Integra Duragen Graft Matrix
Model:ID-1101(1), 1D-1105-(1), 1D-1301-(1), 1D-1305-(1), 1D-2201-(1), 1D-2205-1, 1D-3301(1)
Affected:
Software version:
Recalling Organisation:Integra Neurosciences Pty Ltd (NZ)
12 Florence Avenue
Orewa
Auckland 0931
Contact Information:0800 400 660
Manufacturer:Integra LifeSciences Corporation
Brand Name:Integra Duragen Plus
Model:DP-1013-1, DP-1022-1, DP-1033-1, DP-1045-1, DP-5033
Affected:
Software version:
Recalling Organisation:Integra Neurosciences Pty Ltd (NZ)
12 Florence Avenue
Orewa
Auckland 0931
Contact Information:0800 400 660
Manufacturer:Integra LifeSciences Corporation
Brand Name:Integra Suturable Duragen
Model:DURS2291 DURS3391, DURS4591
Affected:
Software version:
Recalling Organisation:Integra Neurosciences Pty Ltd (NZ)
12 Florence Avenue
Orewa
Auckland 0931
Contact Information:0800 400 660
Manufacturer:Integra LifeSciences Corporation
Brand Name:Integra Neuragen
Model:PNG130, PNG230 , PNG520, PNG620, PNG630
Affected:
Software version:
Recalling Organisation:Integra Neurosciences Pty Ltd (NZ)
12 Florence Avenue
Orewa
Auckland 0931
Contact Information:0800 400 660
Manufacturer:Integra LifeSciences Corporation
Brand Name:Integra Neurawrap
Model:NW320, NH340, NH520, NW740, NW1020, NW1040
Affected:
Software version:
Recalling Organisation:Integra Neurosciences Pty Ltd (NZ)
12 Florence Avenue
Orewa
Auckland 0931
Contact Information:0800 400 660
Manufacturer:Integra LifeSciences Corporation
Issue:Specific lots of some collagen products may have deviated from a production process. The affected lots passed all product finished goods testing, are sterile, and were tested and accepted for endotoxin levels. However, due to a process deviation, they may have been released with higher levels of endotoxins that permitted by release specifications. Higher endotoxin levels may result in fever in the immediate postoperative period. The manufacturer are not aware of any reports of patient injuries or other adverse events in connection with these product lots.
The Health Hazard Evaluation (HHE) provided by the manufacturer details that during an internal quality assurance review of processes, specific lots manufactured during the periods December 2010 to May 2011, and then December 2012 to February 2013, may possibly be affected.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:19/4/2013