Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14511 |
| Brand Name: | Elekta Integrity System |
| Model: | |
| Affected: | Version 1.1 |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Limited (England) |
| Issue: | During beam delivery, the tolerance values are used to do a check for unwanted movements. These values are known as "static tolerances" which are related to all the digital accelerator geometric movement axes except Table Column. The problem is that the "static tolerances" from the calibration files (loaded database) are used for beam delivery instead of the machine calculated values. A fault occurs if the loaded database is incorrect. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 23/4/2013 |





