Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14502 |
| Brand Name: | Bio-Rad Seraclone Anti-S |
| Model: | 808052 |
| Affected: | 2136071 and 22030121 |
| Software version: | |
| Recalling Organisation: | Bio-Rad New Zealand Level 30 48 Shortland Street AUCKLAND 1010 |
| Contact Information: | 0508 805 500 |
| Manufacturer: | Bio-Rad Laboratories Inc |
| Issue: | Latest internal stability data for Seraclone-Anti-S showed a reactivity decrease of the Lot Numbers listed. Weak or false negative results might be possible in testing. Affected batch had only been dstributed to one customer in New Zealand - NZBS. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/4/2013 |