Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:14483
Brand Name:ASP EVOTECH Integrated Endoscope Disinfection System
Model:Part number: 50004
Affected:All serial numbers
Software version:
Recalling Organisation:Johnson & Johnson (New Zealand) Limited
507 Mt Wellington Highway
Mt Wellington
AUCKLAND 1060
Contact Information:0800 922 638
Manufacturer:Advanced Sterilization Products Inc
Issue:The manufacturer has recently determined that under certain circumstances, the EVOTECH Endoscope Cleaner and Reprocessor System may not properly calculate the expiration date of the CIDEZYME XTRA Multi Enzymatic Detergent from the information entered by the user. In these circumstances, the CIDEZYME XTRA Multi Enzymatic Detergent expiration date displayed on the EVOTECH screen is one month after the labeled expiration date.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/4/2013