Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14482 |
| Brand Name: | ASP EVOTECH Integrated Endoscope Disinfection System |
| Model: | Part number: 50004 |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Johnson & Johnson (New Zealand) Limited 507 Mt Wellington Highway Mt Wellington AUCKLAND 1060 |
| Contact Information: | 0800 922 638 |
| Manufacturer: | Advanced Sterilization Products Inc |
| Issue: | The manufacturer has received reports of smoke emanating from the system and upon further investigation has discovered that a circuit board malfunction may occur in the Channel Control Manifold sub-assembly. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 17/4/2013 |