Detail of Recall

Type of Product:Device
Medsafe Reference:14462
Brand Name:Stryker Strykflow Suction Irrigation Tips
Model:Catalogue: 250070403, 250070404, 250070407, 250070408, 250070409, 250070437, 250070483, 250070484, 250070487, 250070488, 250070489, 250070490
Affected:Selected lot numbers
Software version:
Recalling Organisation:Stryker New Zealand Ltd
515 Mount Wellington Highway
Mount Wellington
AUCKLAND 1060
Contact Information:0800 787 9287
Manufacturer:Stryker Endoscopy
Issue:The manufacturer has discovered that the temperatures listed for the pre-vacuum steam sterilization (wrapped method) did not match the validated temperatures for six out of the eleven languages. The pre-vacuum steam sterilization temperature in the Instructions for Use (IFU Rev. K) incorrectly listed 121 - 123°C, instead of the correct 132 - 133°C
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/4/2013