Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14462 |
| Brand Name: | Stryker Strykflow Suction Irrigation Tips |
| Model: | Catalogue: 250070403, 250070404, 250070407, 250070408, 250070409, 250070437, 250070483, 250070484, 250070487, 250070488, 250070489, 250070490 |
| Affected: | Selected lot numbers |
| Software version: | |
| Recalling Organisation: | Stryker New Zealand Ltd 515 Mount Wellington Highway Mount Wellington AUCKLAND 1060 |
| Contact Information: | 0800 787 9287 |
| Manufacturer: | Stryker Endoscopy |
| Issue: | The manufacturer has discovered that the temperatures listed for the pre-vacuum steam sterilization (wrapped method) did not match the validated temperatures for six out of the eleven languages. The pre-vacuum steam sterilization temperature in the Instructions for Use (IFU Rev. K) incorrectly listed 121 - 123°C, instead of the correct 132 - 133°C |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/4/2013 |