Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 14461 |
| Brand Name: | Roche Cobas e602 Module |
| Model: | Cobas 3602 module |
| Affected: | All serial numbers |
| Software version: | |
| Recalling Organisation: | Roche Diagnostics (NZ) Ltd 15 Rakino Way Mt Wellington AUCKLAND 1644 |
| Contact Information: | 0800 363 622 |
| Manufacturer: | Hitachi High Technologies |
| Issue: | The manufacturer advises that under certain conditions a sample will be automatically diluted using an incorrect diluent if there is more than one diluent type onboard at the time. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/4/2013 |